CENTURION Infusion Catheter, 7F, 17cm (DO), Reorder MIC17F7; Infusion Catheter with Antimicrobial Heparin Coating.
FDA Recall Z-1048-2025 · Posted 2025-01-29
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeDYB
Date Initiated2024-12-23
Product Quantity1200 units
Reason for Recall
Certain lots Sterile Infusion Catheters which were mislabeled as having a heparin coating were either distributed or manufactured into medical convenience kits and distributed.
Firm Action
Medline Industries, LP issued one IMMEDIATE ACTION REQUIRED notice and one MEDICAL DEVICE RECALL notice to its affected consignees on 12/23/2024 via email and US Mail. The IMMEDIATE ACTION REQUIRED notice explained the labeling issue with the device, potential risk, and requested the following actions be taken:
1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue corrected labeling to affix to the devices.
The MEDICAL DEVICE RECALL notice, which pertains to kits manufactured with the affected catheters as a component, also explained the labeling issue with the device, potential risk, and requested the following actions be taken: 1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue a label to affix to the devices which will state the affected component shall be removed and discarded from further use.
Root Cause
Process change control
Distribution Pattern
US: FL, KS, ND, PA, WA, WV, IL, WA