Model 509M Pulse Oximetry Interface Module Software versions 2.0 or 2.5
FDA Recall Z-1049-04 · Posted 2004-07-20
Advertisement
Recalling FirmRespironics Novametrix, Inc.
Recall StatusTerminated
Product CodeDQA
Date Initiated2002-09-02
Date Terminated2004-07-06
Product Quantity901 units
Reason for Recall
Audio alarm may fail to sound when SpO2 limits drop below the set limits
Firm Action
Respironics notified customers by letter dated 9/3/02 of the problem with the failure of the audible alarm. A correction would require replacement of the speaker. Non-responders were recontacted by second letter on 2/24/2004.
Root Cause
Other
Distribution Pattern
AK, AZ, CA, CT, FL, NY, NM, PA, OK, TX,