FORUM Archive and Viewer, version 3.1, v 3.1.1, (DVD Format) and v 3.2, v 3.2.1.(DVD and USB Flash Drive Format). Catalog numbers: 000000-20107-750 (DVD with either FORUM v 3.1 or v 3.1.1) 000000-2058-601 (DVD with either FORUM v 3.2 or 3.2.1); 000000-2084-928 (USB drive with FORUM 3.2.1) Ophthalmic image management system.
FDA Recall Z-1049-2015 · Posted 2015-02-04
Advertisement
Recalling FirmCarl Zeiss Meditec AG
Recall StatusTerminated
Product CodeNFJ
Date Initiated2015-01-22
Date Terminated2016-03-04
Product Quantity985
Reason for Recall
Software defect in the FORUM Viewer versions 3.1 and 3.2 which may lead to misinterpretation of the optical coherence tomography (OCT) data.
Firm Action
Important Field Corrective Action for Forum Viewer letters were sent to all affected US customers on January 22, 2015 by certified mail.
Root Cause
Software in the Use Environment
Distribution Pattern
Nationwide Distribution.