Open, Classified

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) OPEN HEART ACCESSORY CDSKit SKU CDS984289O; 2) OPEN HEART CDSKit SKU CDS984355K; 3) BAG A OPEN HEART KITKit SKU DYKM1025D; 4) RR-CABG SUPPLEMENTAL PACKKit SKU DYNJ0394752O; 5) OPEN HEART CDSKit SKU DYNJ902879T; 6) BASIC CARDIAC ANESTHESIAKit SKU DYNJ903523F; 7) OPEN HEART CDSKit SKU DYNJ905557B; 8) AN03 HEART/ANESTHESIA KIT-LFKit SKU PHS972012D.

FDA Recall Z-1050-2026 · Posted 2026-01-12

Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2025-11-26
Product Quantity633 units

Reason for Recall

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

Firm Action

Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution.