Terminated

These models are not available in the US. CONTACT RENEWAL 4 AVT HE (model number M170, M175, M177, M179). Cardiac Resynchronization Therapy Defibrillator. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.

FDA Recall Z-1051-2007 · Posted 2007-07-11

Recall StatusTerminated
Product CodeNIK
Date Initiated2007-03-23
Date Terminated2008-03-02
Product Quantity539

Reason for Recall

Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.

Firm Action

An Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Polynesia, Portugal, Puerto Rico, Qatar, Saint Lucia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. Virgin Islands, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, and West Indies.