Terminated

TAXUS Express Paclitaxel-Eluting MONORAIL Coronary Stent System (3.5 x 32mm). On August 5, 2004 an additional 38 lots were added to this recall, the following TAXUS Express2 Paclitaxel-Eluting Coronary Stent System Part Numbers correspond to the lot numbers listed at the end of the code information below. US part# H7493897024250 H7493897032270 H7493897020300 H7493897020350 H7493897032300 H7493897020350 H7493897024300 H7493897020300 H7493897020300 H7493897032350 H7493897020250 H7493897020270 H7493897024300 H7493897032350 H7493897032270 H7493897020350 H7493897024300 H7493897024250 H7493897024250 H7493897020300 H7493897032350 H7493897032350 H7493897024350 H7493897024250 H7493897024250 H7493897016300 H7493897032350 H7493897024350 H7493897016350 H7493897032300 H7493897016350 H7493897016350 H7493897024350 H7493897016250 H7493897016250 H7493897024300 H7493897024350 H7493897016250 OUS part# H7493800132270 H7493800132270 H7493800132350 H7493800132270 H7493800124350

FDA Recall Z-1052-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeNIQ
Date Initiated2004-07-01
Date Terminated2006-02-22
Product Quantity205,983 units plus additional 3,000 units recalled on 8/5/04.

Reason for Recall

On a few occasions the balloon has failed to deflate within one minute after deployment of the stent.

Firm Action

July 2 and 16, 2004 recall letters were sent to user accounts, such as Hospitals and clinics of the need to recall TAXUS Express2 Paclitaxel-Eluting and Express 2 Coronary Stent Systems. Recall was due to characteristics in the delivery catheters that had the potential to impede balloon deflation during a coronary angioplasty procedure, a condition known as focal neckdown. An additional 38 lots were added to the recall on August 5, 2004, for the same reason.

Root Cause

Other

Distribution Pattern

Nationwide and Canada