Terminated

Lubricated Intubating Stylet 6FR that is inserted into an endotracheal tube to facilitate intubation. This product is used in surgical or medical emergency facilities. Primarily used on infants and pediatric patients.

FDA Recall Z-1054-2015 · Posted 2015-02-04

Recalling FirmCardinal Health
Recall StatusTerminated
Product CodeBSR
Date Initiated2014-12-23
Date Terminated2016-08-01
Product Quantity104,432 stylets

Reason for Recall

A small piece of the stylet sheath has the potential to shear off upon removal of the stylet from the tracheal tube.

Firm Action

URGENT: PRODUCT RECALL letters dated December 23, 2014 were sent to all customers of record. The letters included instructions for customers to: 1) quarantine the affected product; 2) notify any customers whom you may have distributed the affected product to; 3) contact the appropriate Customer Service group to arrange for return and credit of any affected product in your inventory; and, 4) complete and return the enclosed Acknowledgement Form. Customers with questions about this recall can contact Cardinal Health Quality Systems at 800-292-9332.

Root Cause

Process control

Distribution Pattern

US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV, WY) + District of Columbia + Guam + US Army Facility in South Korea