11005BC1 11005BCK1 Flexible Bronchoscope Z21493US-BA (06/2019)

FDA Recall Z-1055-2022 · Posted None

Recalling FirmKarl Storz Endoscopy
Recall StatusOpen, Classified
Product CodeEOQ
Date Initiated2022-04-01
Product QuantityAll Serial Numbers manufactured/distributed since January 2018

Reason for Recall

Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.

Firm Action

On 04/01 and 6/2022, the firm emailed an "Urgent Medical Device Recall" letter to customers and followed up with a hardcopy letter on 04/07-08/2022 via their Third Party contractor informing customers of a labeling update to correct the instructions for use for certain flexible urological endoscopes and advise that high-level disinfection should not be used as a method of reprocessing. Supplemental validation testing of the efficacy of the manual high-level disinfection process using CIDEX OPA was performed. Testing showed that the required efficacy level of disinfection using CIDEX OPA was not achieved. To the extent included in the current applicable instructions for use, the following methods are being removed as reprocessing methods from the instructions for use of the affected endoscopes. +Manual High-Level Disinfection: -Cidex-OPA -Revital-Ox" RESERT (e.g., 2.0% Accelerated Hydrogen Peroxide solution). +Automated Endoscope Reprocessor: -Medivators AER -Reliance EPS -EvoTech ECR Customer Actions: 1. Immediately discontinue the use of all high-level disinfection methods for reprocessing the affected endoscopes. 2. The affected endoscopes should be sterilized after each use by one of the sterilization methods recommended in the instructions for use specific to each endoscope. 3. If your facility does not have access to a sterilization method recommended in the applicable instructions for use, immediately discontinue use of the affected endoscopes and return products to KARL STORZ. 4. Ensure that all relevant personnel in your organization are aware of recall. 5. Updated instructions for use will be available at https://spwebspace.karlstorz.com/sites/HLDCorrection/SitePages/Home.aspx . Once the updated instructions for use are available, discard any prior versions of the instructions for use. Any questions: Telephone: 1-888-352-9616 Fax: 1-888-912-7088 E-Mail: [email protected]

Root Cause

Under Investigation by firm

Distribution Pattern

U.S. Nationwide distribution in the states of FL, KY, LA, MA, MI, MO, MT, NC, NH, NJ, OK, SC, TX and WA.