APD Set with Cassette 3-Prong Peritoneal Dialysis Set Indicated for use in the treatment of patients with renal failure to provide PD fluid exchanges with Baxter HomeChoice PD systems in clinical and home use settings.
FDA Recall Z-1056-2017 · Posted 2017-01-19
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusTerminated
Product CodeFKX
Date Initiated2016-12-16
Date Terminated2017-08-24
Product Quantity8640 eaches
Reason for Recall
Customer complaints received for the presence of leaks
Firm Action
An Urgent Product Recall communication will be sent to affected customers via U.S.P.S., first class mail.
Root Cause
Device Design
Distribution Pattern
US only