Terminated

Brilliance iCT

FDA Recall Z-1058-2017 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-12-23
Date Terminated2018-06-26
Product Quantity1-unit

Reason for Recall

During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.

Firm Action

Philips sent an Customer Information letter. A document showed an incorrect torque wrench used during the assembly process which could potentially apply 4 Newton-meters less torque than specified to the system rotor counterweight bolts. Out of an abundance of caution, Philips will have a Field Service Engineer ensure the bolts on your system are correctly torqued. If you need any further information or support concerning this issue, please contact your local Philips representative. For further information please call (440) 483-5777.

Root Cause

Employee error

Distribution Pattern

Product was not distributed in the US. Unit was distributed in Australia.