Brilliance iCT
FDA Recall Z-1058-2017 · Posted None
Advertisement
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2016-12-23
Date Terminated2018-06-26
Product Quantity1-unit
Reason for Recall
During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
Firm Action
Philips sent an Customer Information letter.
A document showed an incorrect torque wrench used during the assembly process which could potentially apply 4 Newton-meters less torque than specified to the system rotor counterweight bolts. Out of an abundance of caution, Philips will have a Field Service Engineer ensure the bolts on your system are correctly torqued.
If you need any further information or support concerning this issue, please contact your local Philips representative. For further information please call
(440) 483-5777.
Root Cause
Employee error
Distribution Pattern
Product was not distributed in the US. Unit was distributed in Australia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.