ANCILLARY PACK-LF. Sterile Procedural Tray.
FDA Recall Z-1060-2023 · Posted 2023-02-06
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeCAI
Date Initiated2022-11-17
Product Quantity48 cases (96 units)
Reason for Recall
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Firm Action
Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Root Cause
Device Design
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.