Terminated

ACCU-CHEK Inform base unit; catalog no. 3035131.

FDA Recall Z-1061-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeNBW
Date Initiated2004-06-30
Date Terminated2005-10-03

Reason for Recall

Exposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke.

Firm Action

A recall notice dated 6/30/04 was sent to each customer instructing them on proper cleaning methods, asking them to inspect their units for damage, and providing warning stickers for attachment to the base units.

Root Cause

Other

Distribution Pattern

Nationwide.