Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche OPTICCA Cassettes E-Glu. There are 25 cassettes in each box.
FDA Recall Z-1064-04 · Posted None
Advertisement
Recalling FirmOsmetech Inc
Recall StatusTerminated
Product CodeCGA
Date Initiated2004-06-22
Date Terminated2005-01-12
Product Quantity1087 boxes
Reason for Recall
In some cases patient samples with high glucose concentrations, above the reportable range of the OPTI-CCA Analyzer, may be erroneously reported as 'Low'.
Firm Action
Consignees were notified by letter on 6/22/04. A separate Recall Notice was mailed to distributors outside the United States with instructions on issuing the recall notice to their affected customers in the appropriate language.
Root Cause
Other
Distribution Pattern
Nationwide and Argentina, Austria, Bolivia, France, Greece, Germany, Korea, India, Italy, Jordan, Netherlands, Sweden, Switzerland.
Official Source