Terminated

Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 40 cc; Product Number: IAB-05840-LWS. The IAB is utilized for intra-aortic balloon counter-pulsation therapy.

FDA Recall Z-1065-2013 · Posted 2013-04-06

Recall StatusTerminated
Product CodeDSP
Date Initiated2012-11-27
Date Terminated2015-06-02
Product Quantity637 units

Reason for Recall

Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.

Firm Action

Arrow International issued Urgent Medical Device Recall notification on 11/27/12. The letter identified the affected product and the reason for the recall. Accounts are requested to check their stock for the products included within the scope of this recall. Also, customers should cease use and distribution, and quarantine all affected product immediately. Contact Arrow's Customer Service Department at 800-523-8446 for a Return Authorization Number. Customers are to complete and fax the Recall Acknowledgement & Stock Status Form to the number provided, and return the affected product. If customers have questions or need clarification regarding this issue, they should contact Arrow's Customer Service Department at 1-800-523-8446.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- USA, Belgium, Great Britain, Italy, Kazakhstan, Latvia, Russia, Slovenia, Turkey, and South Africa.