Terminated

Philips X-Ray Systems, Allura Xper with R8.2.16 Product Usage: The Allura Xper FD1O and Allura Xper FD1O/10 is intended for use in cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology (EP). The Allura Xper FD2O, Allura Xper FD2O/10 and Allura Xper FD2O/20 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures.

FDA Recall Z-1066-2016 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2015-10-01
Date Terminated2020-10-08
Product Quantity196 devices

Reason for Recall

Upon initiating Fluoroscopy the user may encounter a user message Fluoro failed.

Firm Action

Philips sent a Field Safety Notice to all affected customers on October 1, 2015. The notice informs customers of the issue and the actions planned by Philips. Philips Field Service Engineers are to install Software R8.2.16.1 which addresses the issue. For further information or support concerning the issue, consignees should contact their local Philips Representative or call the Technical Support Line at 800-722-9377.

Root Cause

Software design

Distribution Pattern

US Nationwide in the states of : AK, AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, UT, WI, WV and the country of Canada.