Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
FDA Recall Z-1068-2013 · Posted 2013-04-08
Advertisement
Recalling FirmToshiba American Medical Systems Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2012-11-14
Date Terminated2013-04-08
Product Quantity1 unit
Reason for Recall
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
Firm Action
Toshiba American Medical Systems Inc. sent a Urgent Medical Device Correction letter dated November 14, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customer was informed a New software to prevent the occurrence of this issue will be installed on your system. Toshiba Service Representative will contact you for an appointment to install the New Software on your Aquilion One System.
Please fax the return reply form to (877) 349-3054.
We thank you for your urgent attention to this matter. If you have any further questions regarding this letter please call (800) 521-1968.
Root Cause
Process design
Distribution Pattern
Distributed in the state of MD.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.