Coulter Clone B6-FITC Reagent
FDA Recall Z-1071-03 · Posted 2003-07-30
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Date Initiated2003-04-28
Date Terminated2004-04-20
Product Quantity122
Reason for Recall
Diminished expression on B-Cell populations when drawn in EDTA tubes, which may lead to inaccurate interpretation of phenotype results.
It was determined that choice of anticoagulant may have an effect on results for this reagent. The CD23 clone has shown a diminished expression on B-Cell populations when drawn on EDTA tubes, which may lead to inaccurate interpretation of phenotype results.
Firm Action
Customer letters will be sent to all known purchasers on or about 4/28/2003, also existing inventory will be reworked to include a letter describing performance issues when collection occurs using EDTA.
Root Cause
Other
Distribution Pattern
Nationwide and to Canada
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.