Convoy Advanced Delivery Sheath Kit, Model Numbers: M0045660ST0, M0045661ST0, M0045662ST0, M0045663ST0, M0045665ST0, M0045667ST0, M0045670ST0, M0045672ST0, M0045675ST0, M0045676ST0, M0045690ST0, M0045890ST0, M0045748ST0 and M0045749ST0. Manufactured by Thomas Medical Products, Malvern, PA. Product is intended for use to facilitate the intracardiac placement of interventional devices.
FDA Recall Z-1071-2009 · Posted 2009-04-09
Recalling FirmBoston Scientific Corp
Recall StatusTerminated
Product CodeDYB
Date Initiated2008-12-24
Date Terminated2010-04-06
Product Quantity31,340 of all units
Reason for Recall
The marker band may detach from the sheath and a clinical failure could result in an embolism.
Firm Action
The firm initiated the recall on December 24, 2009 and notified consignees via "Urgent Company Initiated Medical Device Recall - Immediate Action Required" letters which explain the reason for the recall and request the affected product be returned to the Boston Scientific Corporation Distribution Center.
An amendment letter with two additional Convoy Advanced Delivery Sheath Kit models (M0045748ST0 and M0045749ST0) was sent to affected customers on January 27, 2009.
Direct questions about this recall to your local Sales Representative.
Root Cause
Process design
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.