Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Olympus Single Use Guide Sheath Kits with the following product descriptions and model numbers: 1. Single Use Guide Sheath Kit K-201 2.0MM Channel Set: Guide Sheath, Biopsy Forceps, Cytology Brush; Model Number: K-201 2. Single Use Guide Sheath Kit K-202 2.0MM Channel Set: Guide Sheath, Biopsy Forceps; Model Number: K-202 3. Single Use Guide Sheath Kit K-203 2.6MM Channel Set: Guide Sheath, Biopsy Forceps, Cytology Brush; Model Number: K-203 4. Single Use Guide Sheath Kit K-204 2.6MM Channel Set: Guide Sheath, Biopsy Forceps; Model Number: K-204. Used to collect cells or tissue specimens in the respiratory organs.

FDA Recall Z-1071-2025 · Posted 2025-02-11

Recall StatusOpen, Classified
Product CodeEOQ
Date Initiated2025-01-15
Product Quantity454 units

Reason for Recall

Potential for the radiopaque tip of the guide sheath component to disassociate and fall off into the patient.

Firm Action

Olympus notified consignees on about 01/15/2024, via letter. Consignees were instructed to examine inventory and quarantine affected devices, cease usage of the product immediately, acknowledge receipt of the letter through the Olympus web portal, and to forward the notice to all applicable personnel and customers if further distributed.

Root Cause

Under Investigation by firm

Distribution Pattern

US nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑