TunneLoc Tibial Fixation Device (Within an ACL Kit) Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

FDA Recall Z-1072-2018 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeMBI
Date Initiated2017-10-11
Date Terminated2020-05-07

Reason for Recall

Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.

Firm Action

Zimmer Biomet sent an URGENT MEDICAL DEVICE RECALL letter dated October 16, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review inventory and quarantine affected product, which will be returned to Biomet through a sales representative. Questions or concerns can be directed to 574-371-3071

Root Cause

Vendor change control

Distribution Pattern

US Nationwide Distribution