ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Intraflo Flush, Macrodrip (Pole Mount), REF: 42641-06, Sterile EO

FDA Recall Z-1072-2019 · Posted None

Recalling FirmICU Medical, Inc.
Recall StatusTerminated
Product CodeDRS
Date Initiated2018-11-27
Date Terminated2022-02-01

Reason for Recall

Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.

Firm Action

On 11/27/18, Urgent Medical Device Recall notices were sent via traceable mailing to customers informing them to discontinue use and distribution of the affected product. In addition, customers were asked to quarantine and return affected product, and to complete and return response forms. Customers were instructed to inform all potential users of the product within their organization, and if the product had been further distributed, to immediately notify these additional customers. Customers with additional questions were encouraged to call: Global Complaint Management @ 1-844-654-7780 ([email protected]), Medical Information @ 1-800-241-4002 option 6 ([email protected]), or Customer Care @ 1-949-366-4208.

Root Cause

Process design

Distribution Pattern

U.S.: CA, TX, WI, NY, NC, GA, PA, and NY. Foreign: Canada