Single and Dual use Emission Computer Tomography gamma cameras (BrightView XCT, BrightView X & BrightView SPECT) Philips Healthcare Systems, Cleveland, OH. Gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies.

FDA Recall Z-1074-2014 · Posted 2014-02-24

Recall StatusTerminated
Product CodeKPS
Date Initiated2013-12-19
Date Terminated2015-01-12
Product Quantity1,143 units

Reason for Recall

Three problems were reported to the firm: When positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly causing an inappropriate rotation orbit calculation. When the acquisition was started and the detectors began to move into position, the incorrect calculation resulted in the detectors coming in contact with the patient. As

Firm Action

On 12/19/2013 the firm sent Field Safety Notices to their customers.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WA, WI, WV & WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Gabon, Germany, Greece, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom & Viet Nam.