Open, Classified

4 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. M4 Ser CX 311-20 120-240V 50/60Hz, item # 807QS0020CX6.

FDA Recall Z-1074-2022 · Posted None

Recall StatusOpen, Classified
Product CodeFTC
Date Initiated2022-04-26
Product Quantity1

Reason for Recall

Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.

Firm Action

Beginning April 26, 2022 Daavlin Distributing Co. is sent a recall notification letter via certified mail to customers. Customers are to acknowledge receipt of the notification and a company representative will call customers to arrange for return of the devices to update software. If you have any questions or concerns, contact the company at (800) 322-8546. Monday through Friday, 8:00 AM to 4:30 PM, EST.

Root Cause

Software Design Change

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.