Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.
FDA Recall Z-1076-2014 · Posted 2014-02-24
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJIX
Date Initiated2014-01-06
Date Terminated2015-05-06
Product Quantity3496
Reason for Recall
Complaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) and imprecision during PCHE calibration with the Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL).
Firm Action
A recall notification letter, dated January 2014, was sent to Users.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution.