The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.
FDA Recall Z-1076-2015 · Posted 2015-02-10
Advertisement
Recalling FirmMAST Biosurgery USA, Inc
Recall StatusTerminated
Product CodeFTL
Date Initiated2014-08-26
Date Terminated2015-02-24
Product Quantity49 pieces
Reason for Recall
The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July 31, 2017. The correct expiration date was July 31, 2017.
Firm Action
MAST Biosurgery labeled 49 new envelopes and sent them to the distributor, and subsequently had the distributor remove the contents of the outer envelope (instructions for use and the foil pouches containing the product) and place them into the newly supplied and properly labeled outer envelopes provided by them. The distributor requested to send back the original 49 enveloped.
The second batch of new labels was inspected and attached to the new outer envelope to create a properly labeled outer envelope prior to shipping.
The firm requested that once the distributor replaced the product and the IFU into the newly labeled envelopes that they provide documentation that the product had been repackaged into the new envelopes. The distributor was also requested to verify that they did not ship any product to customers with the incorrect label. The firm had the distributor return the 49 improperly with the incorrect labels and the first set of "corrected" labels to them for destruction.
Further questions call (858) 550-8050.
Root Cause
Labeling Change Control
Distribution Pattern
Worldwide Distribution to the country of China.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.