DCI Equipment Track Mount Light Assembly, 110v
FDA Recall Z-1077-2007 · Posted 2007-07-12
Recalling FirmDental Equipment LLC
Recall StatusTerminated
Product CodeEAZ
Date Initiated2007-05-01
Date Terminated2007-07-12
Product Quantity4 dual track lights
Reason for Recall
Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.
Firm Action
On 5/1/07 the dental office contacted the manufacturer via telephone regarding the incident. On 5/2/07 the firm's service representative visited the dental office and replaced the lights.
Root Cause
Other
Distribution Pattern
The lights were distributed to one distributor in NY who in turn distributed the lights to one dental office.