Terminated

ARROW Epidural Needle Component (product number AN-05501). Epidural needles permit access to the epidural space and are used as a conduit for epidural catheter placement.

FDA Recall Z-1080-2015 · Posted 2015-02-10

Recall StatusTerminated
Product CodeCAZ
Date Initiated2015-01-05
Date Terminated2018-07-17
Product Quantity63022

Reason for Recall

The product labeling does not specify the 5 year shelf life of the product.

Firm Action

Urgent Medical Device Recall Notifications, dated January 2015, were sent to direct accounts that identified the product, problem, and actions to be taken. Customers were asked to discontinue use and return affected product.

Root Cause

Labeling design

Distribution Pattern

Distribution US Nationwide, Canada and Mexico.