LH Lithium Heparin tubes, Greiner bio-one, North America, 4238 Capital Drive, Monroe, NC 28110. Venous blood collection tubes.

FDA Recall Z-1081-2012 · Posted 2012-02-22

Recall StatusTerminated
Product CodeJKA
Date Initiated2011-12-14
Date Terminated2012-06-21
Product Quantity2,082,000 pieces

Reason for Recall

The lithium heparin tubes may have a contamination with EDTA potassium.

Firm Action

Greiner Bio-One sent a "RECALL OF LITHIUM HEPARIN TUBES WITH EDTA CONTAMINATION" letter dated December 13, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to complete an attached Product Disposition Site Confirmation Form and to fax the form to 800-726-0052. This form will confirm that customers have discarded/destroyed all affected product. The firm will replace the product after receipt of the form. Contact the firm at 800-515-8112 for questions regarding this recall.

Root Cause

Material/Component Contamination

Distribution Pattern

Nationwide Distribution.