Terminated

Stryker Position Pro Mattress. Mattress, Air Flotation, Alternating Pressure. PositionPRO is a powered pressure relief support surface with low air loss (LAL) intended for medical purposes. PositionPRO consists of multiple air cells filled and emptied by an integrated control unit to provide changes in the distribution of body weight for pressure relief. PositionPRO offers turn assist. PositionPRO is for use with GoBed II, Secure II/3, InTouch, and Epic II 84 in. x 35 in. flat deck frames. This surface is powered via an AC power cord that comes in either a 4 foot long option (PN QDF8087) or 15 foot long option (PN QDF8088). The 4 foot long option is used when the surface is plugged into the auxiliary outlet of the bed. The 15 foot long option is utilized when the surface needs to be plugged into a wall outlet. This product uses a power cord, manufactured by ElectriCord that plugs into the wall outlet or the auxiliary outlet of the bed that the surface is placed onto.

FDA Recall Z-1081-2016 · Posted 2016-03-08

Recall StatusTerminated
Product CodeFNM
Date Initiated2015-12-03
Date Terminated2017-05-24
Product Quantity2694

Reason for Recall

Potential for power cord to melt; hazards include electric shock and fire.

Firm Action

Stryker sent an URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter dated November 10, 2015 to all affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any urgent questions or concerns, please, contact Stryker Customer Service at (1-800-327-0770). Normal business hours are Monday-Friday 8 a.m-5 p.m. (ET).

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Chile, France, Hong Kong, Italy, Mexico, South Korea, Switzerland, United Kingdom.

Official Source

View FDA Recall →