ChemoPlus Drug Spill Kit-INSTI, SKU DP5016K

FDA Recall Z-1081-2021 · Posted None

Recall StatusTerminated
Product CodeLYU
Date Initiated2021-01-07
Date Terminated2024-02-14
Product Quantity137,346 kits

Reason for Recall

Gowns included in the kits potentially have rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns.

Firm Action

URGENT MEDICAL DEVICE CORRECTION NOTICE notification letters dated 1/7/21 were sent to customers. Actions requested on your part: 1) REVIEW your inventory for affected product. 2) QUARANTINE affected product per Attachment 1 within your inventory 3) AFFIX one WARNING LABEL included with this notification to the front of each ChemoPlus" and ChemoBloc" Chemo Spill and Preparation/Administration kits so that it is clearly visible to clinicians, instructing them to remove the affected gown component prior to use. 4) COMMUNICATE this action with all personnel the need to remove the gowns prior to use. 5) NOTIFY any customers to whom you may have distributed or forwarded affected product about this product action. 6) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether or not you have affected product. Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other customers888-444-5440 For questions related to the notification, additional warning labels requests and/or acknowledgement form, please contact the market action team at: [email protected] or call 800-292-9332.

Root Cause

Process change control

Distribution Pattern

Global Distribution. US Nationwide.