Recall StatusOpen, Classified
Date Initiated2022-03-03
Product QuantityOUS: 1,094 units
Reason for Recall
Updated Instructions for Use (IFU) for EXALT Model D
Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.
Firm Action
Boston Scientific notified US and US Territories from March 3rd through March 17th, via hand delivery, FedEx, or e-mail . OUS Consignees were notified from March 3rd through March 10th, via hand delivery, courier, or e-mail. Additional US Consignees located in the US and US Territories were notified from April 7, 2022, through April 13, 2022. Additional OUS Consignees were notified on April 7, 2022.
Immediate Action Required EXALT Model D Single-Use Duodenoscope. Letter states. Customers to whom hand-delivery is not possible within a ten-day timeframe will receive the letter through other means (e.g., courier, e-mail). The notification instructs the account to review the content of the IFU updates, share the information and post it in a visible location near the product, and to complete the Acknowledgment Form to confirm receipt of the notification.
After all applicable regulatory approvals are obtained, the updated IFUs will be
packaged with EXALT Model D Single-Use Duodenoscopes. An updated IFU will be
available at https://www.bostonscientific.com/elabeling in the US.
***UPDATE 4/29/22:
BSC implemented the updated IFU for US customers (an electronic-only IFU published to
BSC's website) on March 21 , 2022. This electronic-only IFU supports UPNs M00542420 and M00542421 (the UPNs intended primarily for US distribution).
BSC will implement the updated IFU for OUS customers (in UPNs M0054242CE0 and
M0054242CE1 , the UPNs intended primarily for OUS distribution) following language
translations and receipt of paper IFU inventory, estimated for May 2022.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium,
Bulgaria, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, IRELAND, Israel, Italy, Kuwait, Netherlands,
Norway, Oman, Poland, Portugal, Qatar, Saudi Arabia, Slovakia, South Africa, Spain,
Sweden, Switzerland, Taiwan, Unit. Arab Emir.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.