WAHL 4212 HEATED MASSAGER, SKU F800000004212 - Product Usage: intended to relieve minor muscle aches and pains.

FDA Recall Z-1084-2021 · Posted 2021-02-05

Recalling FirmWahl Clipper Corp
Recall StatusTerminated
Product CodeISA
Date Initiated2020-12-15
Date Terminated2023-05-09
Product Quantity161 units

Reason for Recall

The connection between the massager and heat element can overheat causing smoke or spark, which may pose a fire hazard

Firm Action

Wahl initiated the recall on 12/15/2020 to its direct consignee by email and posted a notice on its website www.fda.gov on 12/23/2020. The firm is requesting the return of the units.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the states of CT, FL, UT, MO, WI, FL, CA, IL, PA.