NaviCare Nurse Call System provides a comprehensive communication and information system that places patient calls, staff calls, and emergency calls.
FDA Recall Z-1093-2022 · Posted None
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusTerminated
Product CodeILQ
Date Initiated2022-04-13
Date Terminated2023-12-20
Product Quantity11028 units
Reason for Recall
Firm discovered a firmware memory leak with a supplier-manufactured component.
Firm Action
The firm, Hillrom, issued an "Urgent Medical Device Correction Notice", dated 04/13/2022, to impacted consignees via email. The consignee is instructed to identify if they have the impacted product component, Part P2519NNC3A2 and also distribute the notification to personnel as well ensure that the users monitor for system alerts. Notify Hillrom Technical support if there is a loss of connectivity to the nurse call server. Note, error indications will display on multiple user interfaces. Central nurses'
stations will state the room(s) is down (disconnected). In patient room(s), graphical
audio stations will display clear message stating Station down. Contact Nurse
directly . In addition, standard audio stations will have an LED Error state indicator.
Complete and return the response form to [email protected].
Hillrom is testing a firmware update for the Power over Ethernet (PoE) switch and
anticipates availability in April 2022.
Upon receipt of the response form, Hillrom will contact you to schedule an update
to your system once firmware is available.
If you have any questions regarding this Urgent Medical Device Correction, please
contact Hillrom Technical Support at +1 919 854 3361 or [email protected]
Root Cause
Software design
Distribution Pattern
US Distribution to states of: NC and VA.
Official Source