Becton Dickinson's ProbeTec (tm) ET Instrument

FDA Recall Z-1094-03 · Posted 2003-08-20

Recall StatusTerminated
Product CodeLSL
Date Initiated2003-07-21
Date Terminated2003-11-19
Product Quantity1015 units

Reason for Recall

Component of in vitro diagnostic device was incorrectly installed causing incorrect reporting of patient results.

Firm Action

The recalling firm notified consignees by letter on 7/21/2003 and advised of the potential for incorrect optical bundle installation. The firm provided an optical test procedure with materials to conduct an on site performance test. Consignees were instructed not to report patient results until optical performance has been verified. Firm will replace or repair failing instruments.

Root Cause

Other

Distribution Pattern

Product was distributed to domestic and military end users nationwide and international distributors.