Becton Dickinson's ProbeTec (tm) ET Instrument
FDA Recall Z-1094-03 · Posted 2003-08-20
Advertisement
Recalling FirmBecton Dickinson & Co.
Recall StatusTerminated
Product CodeLSL
Date Initiated2003-07-21
Date Terminated2003-11-19
Product Quantity1015 units
Reason for Recall
Component of in vitro diagnostic device was incorrectly installed causing incorrect reporting of patient results.
Firm Action
The recalling firm notified consignees by letter on 7/21/2003 and advised of the potential for incorrect optical bundle installation. The firm provided an optical test procedure with materials to conduct an on site performance test. Consignees were instructed not to report patient results until optical performance has been verified. Firm will replace or repair failing instruments.
Root Cause
Other
Distribution Pattern
Product was distributed to domestic and military end users nationwide and international distributors.