Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Inspira AIR Balloon Dilation System, size 18 x 40; Manufactured by Acclarent, Inc. 1525-G O'Brien Drive, Menlo Park, CA 94025 Product Usage: Dilation of airway tree.

FDA Recall Z-1095-2012 · Posted 2012-03-01

Recalling FirmAcclarent, Inc.
Recall StatusTerminated
Product CodeKTI
Date Initiated2012-01-30
Date Terminated2013-08-29
Product Quantity633 units

Reason for Recall

Acclarent received reports of difficulty deflating the balloon during the procedure, which could potentially result in airway obstruction.

Firm Action

Acclarent sent an "Urgent Voluntary Product Recall" letter dated January 31, 2012 by US mail, return receipt requested to all affected customers. Additionally, customers known to have scheduled procedures using the affected product were notified by phone to advise them of the recall. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory immediately, discontinue use and return any remaining in stock per the instructions provided. The letter provides additional information and photographs illustrating how to identify the recalled product. Customers were advised to complete the enclosed Business Reply Card regardless of whether they have or do not have product subject to this recall in their possession and mail to Stericycle. Customers were instructed to distribute this information to all staff within their department who uses the affected product. For further questions call 1-866-781-1173 or contact your Acclarent Sales Representative.

Root Cause

Component design/selection

Distribution Pattern

Nationwide Distribution (USA) only sixe 18 x 40 Worldwide distribution other sizes.Distirbution to France, Canada, Saudi Arabia

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑