Recall StatusOpen, Classified
Date Initiated2025-01-08
Product Quantity1351 units
Reason for Recall
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN for the TLiIQ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
Firm Action
Hologic issued a "Urgent: Medical Device Recall Notification on January 8, 2025 via FedEx. Hologic asked consignees to take the following actions:
1. Please discontinue use of the Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 2. Please segregate and count how many of the affected Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 3. Please fill out the Customer Response Form (CRF); 4. Hologic will process and ship the replacement kits as listed on the completed CRF once it has been received; 5. For tests already administered from the affected product lots and if the patient has not given birth, it is at the
discretion of the healthcare providers to notify patients who were managed based on a number of clinical decision
points including but not exclusive to fetal fibronectin test results. If the patient has given birth, no further action is
required.; 6. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Guam and the countries of Canada and Japan.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.