Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

DataLink 2000 Data Manager

FDA Recall Z-1096-04 · Posted 2004-07-28

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Date Initiated2003-11-03
Date Terminated2004-08-30
Product Quantity361

Reason for Recall

Software anomaly allows results to be archived prior to validation or to be uploaded to the lab information system.

Root Cause

Other

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑