The SYMBIONIC LEG is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. Product Usage: SYMBIONIC LEG is intended for continuous use by amputees of low to moderate impact levels that weigh less than 125 kg for foot categories 25 30 and less than 100kg for foot categories 22, 23 and 24. The system is not intended for athletic use.
FDA Recall Z-1097-2014 · Posted 2014-02-26
Advertisement
Recalling FirmOssur Americas, Inc.
Recall StatusTerminated
Date Initiated2012-10-15
Date Terminated2014-02-28
Product Quantity144 units
Reason for Recall
OSSUR is recalling the Symbionic Leg because it is not recognizing the low battery warning signal from the Symbionic Leg before it powers down.
Firm Action
Ossur Americas contacted all affected customers via phone on October 15, 2012 informing them that a software update for this product is required, along with the recommended instructions to return the product. A loaner prosthetic will be provided to the patient while the software update is being implemented.
Root Cause
Other
Distribution Pattern
US Nationwide distribution.