Terminated

Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).

FDA Recall Z-1098-2016 · Posted 2016-03-11

Recalling FirmDeerfield Imaging
Recall StatusTerminated
Product CodeLNH
Date Initiated2016-02-10
Date Terminated2017-04-05
Product Quantity35 USA, 14 OUS

Reason for Recall

Emergency helium venting lines (i.e. quench lines) may not have been properly installed.

Firm Action

Consignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or +1-866-475-0525.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.