Softbank software Product Usage: Decision support software for transfusion service. It keeps track of inventory from outside sources, multi-site inventory control, records of testing of units, and allows for record keeping for transfusion preparation.

FDA Recall Z-1098-2017 · Posted None

Recall StatusTerminated
Product CodeNSX
Date Initiated2016-10-06
Date Terminated2019-08-21
Product Quantity189 total

Reason for Recall

Software error. Potential for incorrect results

Firm Action

Consignees were sent a recall notification, dated 10/06/2016, providing instructions for software correction. The firm indicates that the issue will be resolved with a mandatory hot fix. Hot fix will be released when its effectiveness is confirmed. Consignees are instructed to acknowledge receipt of the recall notification and grant permission for the recalling firm to load the hot fix when available. The firm can be contacted by calling Technical Support at 1-800-763-8522. Until the hot fix is available, consignees can still operate the software by exiting (Esc-Quit) the Test Verify option after each save changes (F12-Accept).

Root Cause

Software design (manufacturing process)

Distribution Pattern

Worldwide Distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, WI, WV, and the countries of Hawaii and Puerto Rico Canada & Jamaica