Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090

FDA Recall Z-1098-2026 · Posted 2026-01-15

Recalling FirmMedtronic Inc
Recall StatusOpen, Classified
Product CodeQAN
Date Initiated2025-12-04
Product Quantity442

Reason for Recall

Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

Firm Action

Beginning 04-Dec-2025, Medtronic distributed an Urgent Medical Device Recall notice to all U.S. consignees who have impacted products via mail courier. Consignees were instructed to complete and return a response form to the firm and to return any affected product they may have.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.