Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

FDA Recall Z-1100-2026 · Posted 2026-01-15

Recalling FirmICU Medical Inc.
Recall StatusOpen, Classified
Product CodeJKA
Date Initiated2025-12-19
Product Quantity64290 total

Reason for Recall

Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

Firm Action

On December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.