Terminated

Baxter Interlink System Minivolume Extension Set, product code 1C8483; 18'', 1.3 mL Vol., with 0.22 Micron Downstream Filter, 2 injection sites and a male Luer Lock adapter; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

FDA Recall Z-1101-05 · Posted 2005-08-06

Recall StatusTerminated
Product CodeFPA
Date Initiated2005-07-06
Date Terminated2007-01-31
Product Quantity3,072 sets

Reason for Recall

Leaks and disconnection between the bifurcated Y-site and the set tubing may occur and result in underdelivery of medication, blood loss, and a breach of the sterile fluid pathway.

Firm Action

Baxter sent Urgent Product Recall letters dated 7/6/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the complaints of leaks and disconnection between the bifurcated Y-site and the set tubing, resulting in the possible under delivery of medication, blood loss and a breach of the sterile fluid pathway. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837.

Root Cause

Other

Distribution Pattern

Arizona, Georgia, Illinois, Michigan, Ohio, Pennsylvania, South Carolina and Texas.