Recall StatusTerminated
Date Initiated2012-10-15
Date Terminated2013-11-06
Product Quantity27 units
Reason for Recall
It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may
become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system
is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to
operate normally, and the risk of a serious health hazard is therefore very low. The LCD Moni
Firm Action
Toshiba sent an Urgent Medical Device Correction letter dated October 1, 2012 via USPS return receipt mail to all affected customers . The letter identified the affected product, problem and actions to be taken. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions contact your local Toshiba representative at 800-521-1968.
Root Cause
Nonconforming Material/Component
Distribution Pattern
USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT.