Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.
FDA Recall Z-1102-06 · Posted 2006-06-10
Advertisement
Recalling FirmBoston Scientific
Recall StatusTerminated
Product CodeDQY
Date Initiated2006-05-22
Date Terminated2006-12-30
Product Quantity75
Reason for Recall
Gaps were identified between polymer segments on the 7F Wiseguide Guide Catheter in which the PTFE was not properly laminated to the braid of the catheter. This delamination of the PTFE from the braid could compromise the device delivery.
Firm Action
Affected customers were mailed an Urgent Medical Device Recall or Correction - Immediate Action Required letter, dated 05/22/2006. The letter describes the issue, states to immediately discontinue use of and segregate recalled product, provides instructions to ship product back to Boston Scientific and asks for the Reply Verification Tracking Form to be faxed to Boston Scientific in acknowledgement.
Root Cause
Other
Distribution Pattern
Nationwide- States of AZ, CA, CO, FL, HI, KS, MI, NV, OK, TX, VA.