DELFIA PC with V3.0 Software, Fluorometer, for clinical use Product Number: 1235-8100 Product Usage: The Wallac 1235 AutoDELFIA automatic immunoassay system is designed to automatically perform assays using the DELFIA technology. DELFIA technology is based on the proven and widely used method of time resolved fluorometry

FDA Recall Z-1102-2017 · Posted None

Recall StatusTerminated
Product CodeKHO
Date Initiated2016-12-21
Date Terminated2018-10-03
Product Quantity1 US ; 3 OUS

Reason for Recall

AutoDELFIA System with Version 3.0 Software: Potential plate loading failures

Firm Action

PerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software version 3.0 revision 1 or 2 will be updated through the correction to revision 3 or the latest released revision

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,