WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0
FDA Recall Z-1102-2023 · Posted 2023-02-07
Advertisement
Recalling FirmWishbone Medical, Inc.
Recall StatusOpen, Classified
Product CodeKTT
Date Initiated2023-01-04
Product Quantity195
Reason for Recall
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
Firm Action
The firm visited the affected consignee on December 30, 2021, to discuss the product issue.
Root Cause
Process design
Distribution Pattern
US Nationwide distribution in the state of South Carolina.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.