Terminated

AIRIS II, Altaire Nuclear Magnetic Resonance Imaging Device Systems

FDA Recall Z-1104-06 · Posted 2006-06-10

Recall StatusTerminated
Product CodeLNH
Date Initiated2006-04-28
Date Terminated2008-11-24
Product Quantity5

Reason for Recall

Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.

Firm Action

The firm''s Field Service personnel will apprise the customers of this situation during scheduled visits at each customer location. During this visit, the corrected software will be installed by the Hitachi Field Service Technician. The installation of the software upgrades by the service technicians began on 4/28/2006. The firm anticipates that the correction of all units in their inventory as well as those units installed out-in-the-field will becompleted by 6/15/2006.

Root Cause

Other

Distribution Pattern

The devices were distributed to five (5) customers located in four (4)states: Colorado, Ohio, Nebraska and Texas