Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

FDA Recall Z-1105-2007 · Posted 2007-08-02

Recall StatusTerminated
Product CodeDWQ
Date Initiated2004-10-12
Date Terminated2007-08-02
Product Quantity274 units.

Reason for Recall

Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.

Firm Action

Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.

Root Cause

Other

Distribution Pattern

Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.