Terminated

LIFEPAK 12 defibrillator/monitor Part Numbers: VLP12-02-000000 through VLP12-02-005003, AVLP12-02-000000 through AVLP12-02-005003, DVLP12-02-000000 through DVLP12-02-005003, EVLP12-02-000000 through EVLP12-02-005003, FVLP12-02-000000 through FVLP12-02-005003, LVLP12-02-000000 through LVLP12-02-005003, PVLP12-02-000000 through PVLP12-02-005003, UVLP12-02-000000 through UVLP12-02-005003

FDA Recall Z-1107-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeMKJ
Date Initiated2004-05-05
Date Terminated2006-05-06
Product Quantity20,676 domestic, 19,899 international

Reason for Recall

Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in intermittent ECG signal; therapy connector has potential for impact damage.

Firm Action

The firm reviewed records and/or contacted customers by telephone to verify affected devices used in the out of hospital mobile environment. All affected customers were notified by a letter dated May 2004. The firm''s service representative will contact customers to inspect and correct the devices.

Root Cause

Other

Distribution Pattern

The firm distributes to distributors, hospitals, EMT-Ds and paramedics nationwide. The firm also distibutes the product internationally.